TENS-2000
3 Mode without Timer
Channels: Dual, Isolated between channels
Modes: Burst, Constant, Modulation
Pulse Width: 60 to 250 microseconds, adjustable
Pulse Amplitude (intensity): 0 to 80 mA each
channel, adjustable (500 ohm load)
Pulse Frequency: 2 to 150 Hz adjustable
Modulation Mode: Pulse width automatically varies is
a cyclical pattern over an interval of nominally 4 seconds
Burst Mode: 7 pulses per burst, 2 bursts per second,
100Hz internal frequency of burst
Battery Life: 70 hour average (alkaline)
Waveform: Modified square wave with zero net direct
current
|
|

|
TENS-2500
3 Mode with Timer
Channels: Dual, Isolated between channels
Modes: Burst, Constant, Modulation
Pulse Width: 60 to 250 microseconds, adjustable
Pulse Amplitude (intensity): 0 to 80 mA each
channel, adjustable (500 ohm load)
Pulse Frequency: 2 to 150 Hz adjustable
Timer: 30 Min, 60 Min, or Constant
Modulation Mode: Pulse width automatically varies is
a cyclical pattern over an interval of nominally 4 seconds
Burst Mode: 7 pulses per burst, 2 bursts per second,
100Hz internal frequency of burst
Battery Life: 70 hour average (alkaline)
Waveform: Modified square wave with zero net direct current
|
|

|
TENS-6000
Digital TENS with Timer
Channels: Dual
Modes: Burst, Normal, MRW, SD, Bi-Pulse
Pulse Width: 50-300 microseconds, adjustable
Pulse Amplitude (intensity): 0 to 80 mA each
channel
Pulse Frequency: Variable, 2-150 Hz
Waveform: Asymmetric biphasic square pulse
Timer: 5 to 90 Min, or constant time
Compliance Timer: 60 sets of operation records,
999 hours record time
|
 |
Ultimate Ortho TEMS Digital TENS/EMS Combination Unit
Channels: Dual
Output: 5mA to 100mA into 1K_
Pulse Width: 50_S to 400_S adjustable
Pulse Rate: 1Hz to 150Hz adjustable
Waveform: Symmetrical Bi-phasic rectangular
Symmetrical Alternated rectangular
Treatment Timer: Continuous, 15min, 30min,
45min, 60min, 90min |
|

|
EMS-3000
Muscle Stimulator
Channels: Dual, Isolated between channels
Modes: Burst, Constant, Modulation
Pulse Amplitude: Constant current 0 to 80 mA each
channel, adjustable (500 ohm load)
Pulse Frequency: Adjustable, 5, 30, 100 Hz
Cycle On/Off: 1-30 sec.
On Ramp: Adjustable, 1, 3, 5 seconds
Battery Life: Approx. 70 hours
Waveform: Modified square wave with zero net
direct current
|
|

|
EMS-6500
Digital EMS with Timer
Channels: Dual
Modes: Synchronous (S), Constant (C), and Alternate (A)
Pulse Amplitude: 0 to 80 mA peak into 500 ohm load
each channel, adjustable
Pulse Frequency: 2 to 120 Hz
Pulse Ramp: 1 to 8 seconds
Contraction: Variable Control (2-99 Sec)
Relaxation: Variable Control (2-99 Sec)
Waveform: Asymmetric biphasic square pulse wave
with zero net direct current
Compliance Timer: 60 sets of operation records,
999 hours record time
|
 |
Combo Unit
TENS/EMS
The Combo T/E Unit is battery operated pulse generator that sends electrical impulses electrodes to the body and reach the nerves and underlying muscle group. This unit is a combination stimulator of TENS and EMS which can be used for muscle stimulation and pain relief. The device is provided with two controllable output channels, each independent of each other. An electrode pair can be connected to each output channel. The intensity controls are protected by a cap to avoid accidental touch. The settings are controlled by press buttons. |
 |
IF-2206A
Digital “Piggy Back”
Channels: Dual, isolated
Pulse Amplitude: Adjustable, 0-60 mA into 500ohm
each channel
Carrier Frequency: 4000Hz fixed (channel 1)
Modulating Frequency: 4001-4160 adjustable
(channel 2)
Constant Mode: 4-160 bps, adjustable
Auto Sweep: 8-145 bps, 4-45 bps, 4-set bps
Frequency Shift Percent: Shifts from 30% below set
frequency to 60% above then back 30% below set frequency
Output Configure: Quadra polar (4 electrodes)
Wave Form: Symmetrical Balanced Sine Wave
Interference Pulse Frequency: 4-160 bps; adjustable
to 4 bps/step
Pulse Duration: 125 s maximum
Patient Compliance Meter: Displays treatment times
Pateint Lock System: Prevents patient from changing
the parameters set by the physician/clinician
Timer: 15 and 30 minutes |
|

|
IF-2206
Digital True Sine Wave
Channels: Dual, Isolated Between Channels
Pulse Amplitude: Adjustable, 0-60 mA into 500 Ohm
each channel
Carrier Frequency: 4000 Hz fixed (Ch. 1)
Modulating Frequency: 4001-4160 Adjustable (Ch. 2)
Frequency Function Mode: Constant Mode (C): 4-160
bps; Adjustable
Auto Sweep (A): 80-145 bps, 4-45 bps, 4-SET bps
Frequency Shift (S): 1/1 Abruptly Shift, 6/6 Abruptly
Shift, 6/6 Ramped,10/10 Abruptly Shift, 10/10 Ramped
Sweep Time: 15 seconds
Frequency Shift Percent: Shifts from 30% below set
frequency to 60% above then back to 30% below set
frequency.
Output Configuration: Quadra polar (4 electrodes)
Wave Form: Symmetrical Balanced Sine Wave
Interference Beat Frequency: 4-160 bps; adjustable
to 4 bps/step
|
|

|
IF-4250
Interferential Stimulator
Channel: Dual
Pulse Width: 125 microseconds for each phase
Pulse Amplitude: Adjustable 0-50 mA peak
at 500 S/channel
Carrier Frequency: 4000 Hz fixed
Difference Frequency: (1 +/- 0.5) ~
(250+/-10%) Hz adjustable
Interference Frequency: Channel B
Auto Sweep: 4-45 bps, 80-150 bps,
4-set
Constant Mode: 4-400 bps, adjustable
Waveform: Symmetrical Biphasic Square
Four Frequency Shift: Continuous, 1/1
abrupt, 8/8 abrupt, 10/10 ramped
|
|

|
PGS-123
Pulsed-Galvanic Stimulator
Channels: Single
Output Mode: Synchronous or Alternation.
Alternation mode-LED alternates between green
and red color
Pulse Rate: 2-100 Hz
Wave Form: Positive or negative wave form
Pulse Duration: 5 s
IntraPulse Interval: 100 s.
|
|

|
Micro-2000
Analog Microcurrent
Channels: Dual
Output Waveform: Modified square DC biphasic pulses of 0.3Hz, 8Hz, and 80Hz changing polarity at intervals of 1 seconds.
Timer: 20, 40 minutes and constant
Frequency: 0.3, 8 and 80Hz
Output Voltage: 12 Volts
|
|